31/10/2024
The HERC line meets medical standards for worldwide use up to 5000 m. Units have 2 x MOPP protection for patient and user safety and <100 μA leakage current. Compliant with IEC 62368-1, it's recommended for industry, IT, and audio/video. EMC meets IEC 60601-1-2 for medical devices, and devices can also be used domestically according to IEC 60601-1-11. Units > 75 W have active PFC for permissible harmonic current values.
These medically certified power supplies combine a compact design with high efficiency. The devices also offer optimised heat management and flexible integration, making them ideal for a wide range of applications.
- HERC18: Output voltage: 5V / 12V / 15V / 24V
- HERC30: Output voltage: 5V / 12V / 15V / 24V / 48V
- HERC60: Output voltage: 12V / 24V / 48V
- HERC175: Output voltage: 12V / 24V / 36V / 48V
- HERC250: Output voltage: 12V / 24V / 48V